与计算机连接的医疗设备安全问题
Security Problems in the Use of Medical Equipment Connected with
Computer
摘要 阐述了医用电气设备与计算机及其类似设备连接使用的方式。医用电气设备与计算机这两个设备本身检测标准不同,对漏电流的要求也不相同,在利用医用设备进行治疗时,计算机所产生的漏电流存在严重的安全隐患。针对这种情况展开分析,以期企业在生产类似医疗设备时能引起足够的重视,消除隐患。
关键词 漏电流; 计算机; 医用电气设备
Abstract Describes the connect use of medical electrical equipment with computer or
other similar equipments. As medical electrical equipment and computer equipment
have different testing standards and different requirements to leakage current, the
leakage current generated from computers will be a serious hazard to patients during
treatment with medical equipment. Through analysis of this situation,this paper tries to
arouse manufacturers' attention to the hazard, so as to eliminate possible sources of
danger during their production of this kind of medical equipment.
Keywords leakage current ;computer ;medical electrical equipment